2.2
Test
002:
Testing
the
Gas
Supply
This
test
allows
you
to
test
the
gas
supply.
Press
the
Reset
key
to
select
Test
002.
Press
the
Info
key
and
the
Reset
key
simultaneously
to
activate
Test
002.
Display:
Pressure
high
The
gas
supply
pressure
is
sufficient.
Disconnect
the
Oxylog
2000
from
the
gas
supply.
Set
the
NoAirMix/AirMix
mode
selector
switch
to
the
centre
position
in
order
to
vent
the
residual
pressure.
Display:
The
gas
supply
pressure
is
too
low
or
the
device
is
not
connected
to
a
gas
supply
source.
Pressure
low
If
problems
occur
during
this
test,
first
calibrate
the
pressure
sensors
(Test
11
in
the
Extended
Service
Mode
ESM)
Press
the
Info
key
and
the
Reset
key
simultaneously
to
return
to
the
Customer
Service
Mode
(CSM).
Exiting
the
Customer
Service
Mode
(CSM)
Switch
the
Oxylog
2000
off
and
on
again
to
return
to
the
Oxylog
2000
normal
operating
mode.
DS-Modus
5503.160
Oxylog
2000
11/97
Service-Mode
SW
2.xx
Page
11

Table of Contents

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals