Dräger Oxylog 2000 - General Information about the Drager Service Mode SW; Definition of the Original Status

Dräger Oxylog 2000
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
For
internal
use
only.
Copyright
reserved.
Drager
Medizintechnik
Drager
1
General
Information
About
the
Drager
Service
Mode
SW
2.xx
You
can
operate
the
DragerService
Mode
with
the
Info
key
and
the
Reset
key.
These
keys
allow
you
to
change
the
set
values,
to
scroll
through
tests,
or
to
confirm
execution
of
a
specific
test.
The
DragerService
Mode
is
subdivided
into
the
Customer
Service
Mode
(CSM)
and
the
Extended
Service
Mode
(ESM).
The
working
language
used
throughout
the
DragerService
Mode
is
English.
The
DragerService
Mode
in
Oxylog
2000
has
17
tests:
The
first
five
tests
(Test
001
through
Test
005)
are
specifically
designed
for
the
customer
(CSM),
but
can
also
be
used
by
Drager
service
technicians.
The
other
twelve
tests
(Test
011
through
Test
022)
are
part
of
the
Extended
Service
Mode
(ESM)
which
is
used
to
carry
out
calibrations.
These
tests
are
only
accessible
to
Drager
service
technicians.
To
do
so,
the
Drager
service
technician
must
enter
a
numeric
code.
1.1
Definition
of
the
Original
Status
Each
function
in
the
DragerService
Mode
can
be
carried
out
separately.
At
the
beginning
of
the
DragerService
Mode
and
after
completion
of
a
specific
test,
the
Oxylog
2000
enters
one
of
the
following
original
status.
Table
1:
Original
Status
of
Oxylog
2000
in
DragerService
Mode
l:E
solenoid
valve
off
(expiration)
PN
solenoid
valve
off
(PEEP
is
inactive)
autozero
solenoid
valve
operates
automatically
MV
valve
closed
alarm
LED
off
buzzer
off
LCD
backlight
on
alarms
deactivated
DS-Modus
5503.160
Oxylog
2000
11/97
Service-Mode
SW
2.xx
Page
3