2.4
Test
004:
Buzzer
Test,
LED
Test,
and
LCD
Test
This
test
allows
you
to
check
the
audible
and
visible
signals.
The
LEDs
of
the
internal/external
power
supply,
the
alarm
LED
and
the
backlight
of
the
display
come
on.
Press
the
Reset
key
to
select
Test
004.
Press
the
Info
key
and
the
Reset
key
simultaneously
to
activate
Test
004.
Display:
LCD_HG
Test
ON
RESET
Test
OFF
Press
the
Info
key
to
start
Test
004.
The
buzzer
is
activated
and
the
pixels
of
the
display
are
triggered.
Press
the
Reset
key
to
stop
Test
004.
Press
the
Info
key
and
the
Reset
key
simultaneously
to
return
to
the
Customer
Service
Mode
(CSM).
Exiting
the
Customer
Service
Mode
(CSM)
Switch
the
Oxylog
2000
off
and
on
again
to
return
to
the
Oxylog
2000
normal
operating
mode.
DS-Modus
5503.160
Oxylog
2000
11/97
Service-Mode
SW
2.xx
Page
13