EasyManua.ls Logo

Dräger Oxylog 2000 - Positioning and Gas Supply Connection

Dräger Oxylog 2000
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
For
internal
use
only.
Copyright
reserved.
Drager
Medizintechnik
Drager
3.2.2
Software
Version
3.xx
Display:
Set
Pmax
=
60
mbar.
0
mbar
pressure
Set
the
mode
switch
for
Pmax
to
60
mbar.
Press
the
Info
key
and
the
Reset
key
simultaneously
to
start
the
calibration
mode.
NoAirMix
xx
L/min
ENTER
=
calibration
point
Press
the
Info
key
or
the
Reset
key
to
set
the
flow
so
that
the
value
shown
on
the
display
screen
matches
the
value
measured
on
the
external
flowmeter.
Press
the
Info
key
and
the
Reset
key
simultaneously
to
save
the
following
calibration
values
in
the
EEPROM.
Calibration
Value
Tolerances
for
Set
Flow
Values
10
9.5-10.5
L/min
20
19.0-21.0
L/min
30
29.0-30.0
L/min
50
49.0-51.0
L/min
60
59.0
-
60.0
L/min
The
last
calibration
value
is
calibrated.
Oxylog
2000
automatically
returns
to
the
Extended
Service
Mode
(ESM).
Exiting
the
Extended
Service
Mode
(ESM)
If
it
is
switched
off
and
then
on
again,
Oxylog
2000
returns
to
normal
operation.
DS-Modus
5503.160
Oxylog
2000
11/97
Service-Mode
SW
2.xx
Page
20

Table of Contents

Other manuals for Dräger Oxylog 2000

Related product manuals