1.4
Error
Messages
During
a
Specific
Test
If
the
system
detects
and
displays
an
error,
you
must
confirm
this
error
by
pressing
the
Info
key
and
the
Reset
key
simultaneously.
Once
the
error
has
been
removed,
specific
tests
can
be
selected
and
carried
out
again.
All
potential
errors
and
their
causes
are
listed
below:
1.4.1
Power
Supply
Low
Problem:
battery
empty
No
external
power
supply
connected.
The
internal
power
supply
is
no
longer
sufficient
for
safe
operation.
1.4.2
P
ressure
Low
Problem:
pressure
low
Gas
supply
pressure
is
too
low.
1.4.3
P
ressure
High
Problem:
pressure
high
A
gas
supply
source
is
connected
although
no
pressure
is
required
for
this
test.
1.4.4
M
o
de
Switch
Position
(AirMix/NoAirMix)
Problem:
NoAirMix
!
The
mode
switch
is
not
in
the
correct
position.
or
Problem:
AirMix
The
mode
switch
is
not
in
the
correct
position.
DS-Modus
5503.160
Oxylog
2000
11/97
Service-Mode
SW
2.xx
Page