For
internal
use
only.
Copyright
reserved.
Drager
Medizintechnik
3.11
Test
021:
Testing
the
Minute
Volume
Valve
(Fast)
During
this
test,
the
minute
volume
valve
control
is
switched
off.
The
minute
volume
valve
can
be
opened
or
closed
fast
by
hand.
A
Caution:
The
minute
volume
valve
might
get
stuck
in
the
mechanical
stop.
To
avoid
this,
do
not
move
the
minute
volume
valve
completely
to
its
mechanical
stops.
Open
or
close
the
minute
volume
valve
carefully.
Select
the
Extended
Service
Mode
(ESM).
Press
the
Reset
key
or
the
Info
key
to
select
Test
021.
Press
the
Info
key
and
the
Reset
key
simultaneously
to
activate
Test
021.
Display:
LCD_HG
MV
close
RESET
MV
open
Press
the
Info
key
to
fast
close
the
minute
volume
valve.
Press
the
Reset
key
to
fast
open
the
minute
volume
valve.
Press
the
Info
key
and
the
Reset
key
simultaneously
to
return
to
the
Extended
Service
Mode
(ESM).
Exiting
the
Extended
Service
Mode
(ESM).
Switch
the
Oxylog
2000
off
and
on
again
to
return
to
the
Oxylog
2000
normal
operating
mode.
DS-Modus
5503.160
Oxylog
2000
11/97
Service-Mode
SW
2.xx
Page
37

Table of Contents

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals