ecom-ST 1.0 35
The status of the device can also be queried via the analog outputs (31 Stat
:). It is only possible to assess whether the measured values are valid in
connection with the status of the device. The table below lists all possible
statuses.
Analog value
Meaning
4 mA
Device error
6 mA
Calibration phase not finished
7 mA
Device on standby
8 mA
Sample gas pump off
9 mA
Fresh air purging active
10 mA
Device in measuring mode, but the CO sensor is
switched off
20 mA
Device in measuring mode
HINWEIS
Valid measured values are only available if the status is 20 mA.