EasyManua.ls Logo

Ecom ST - Technical Data; Measured values

Ecom ST
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 ecom-ST 1.0
Technical Data
EC = Electrochemical sensor
IR = Infrared sensor
K = Catalytic Sensor
T = Thermocouple NiCr-Ni
PT = PT 2000
D = DMS bridge
MV* = Measured value (higher value prevails)
Measured values
Range
Accuracy
O
2
EC
0 vol. % − 21 vol. %
±0,3 vol. %
CO
EC
0 ppm − 4000 ppm
±20 ppm / 5 % MV*
CO H
2
- comp.
EC
0 ppm − 10.000 ppm
±20 ppm / 5 % MV*
CO%
EC
0 ppm − 63.000 ppm
±100 ppm / 5 % MV*
NO
EC
0 ppm − 5000 ppm
±5 ppm / 5 % MV*
NO ExtraLow
EC
0 ppm − 300 ppm
±2 ppm / 5 % MV*
NO
2
EC
0 ppm − 1000 ppm
±5 ppm / 5 % MV*
NO
2
Low
EC
0 ppm − 100 ppm
±5 ppm / 5 % MV*
SO
2
EC
0 ppm − 5000 ppm
±10 ppm / 5 % MV*
SO
2
Low
EC
0 ppm − 100 ppm
±5 ppm / 5 % MV*
H
2
EC
0 ppm − 20.000 ppm
±100 ppm / 5 % MV*
H
2
S
EC
±10 ppm / 5 % MV*
CO
2
IR
0 vol.% − 20 vol.%
±0,5 vol. % / 5 % MV*
CO
2
IR
0 vol.% − 100 vol.%
±5 vol. %
CH
4
IR
0 vol.% − 5 vol.%
±0,2 vol. % / 5 % MV*
C
x
H
y
K
0 vol.% − 4 vol.%
T-Gas
T
0 °C − 500 °C
±2 °C / 1,5 % MV*
T-Air
PT
0 °C − 99 °C
±1 °C
pressure
D
0 hPa − ± 100 hPa
±0,5 hPa / 1 % MV*
Air pressure
D
300 hPa − 1100 hPa
±2 hPa