H100B Pulse Oximeter User Manual
- 44 -
system for seconds (the pause period can be set to 30, 60,
90 or 120s by the user, see section 3.3.5), the audio
alarm off icon displays. But the visual alarm is still
active. For example, if the measured SpO
2
value is
higher than SpO
2
Hi Alarm or lower than SpO
2
Lo
Alarm, there will beoricon displayed on screen, and
the SpO
2
or PR character will flash.
If Alarm system is set to OFF, all audio alarms and
visual alarms are turned off.
WARNING
When the Alarm system is off, the oximeter will not
give an alarm prompt. In order to avoid endangering
the patient’s life, the user should use this function
cautiously.
3.3.10 SpO
2
Alarm Setup
The user can choose SpO
2
Hi Alarm and SpO
2
Lo
Alarm in menu to adjust SpO
2
alarm limit. Press the Up
Arrow button or Down Arrow button to increase or
decrease alarm limit.
By default, SpO
2
Hi Alarm and SpO
2
Lo Alarm in Neo