H100B Pulse Oximeter User Manual
- 76 -
Capacity 1500 mAh/1800 mAh
Typical battery life 30 hours/36 hours
Charge time 2.5 hours to 80%
4 hours to 100%
Note: Only the battery of 1500mAh capacity meets
the UL requirements.
A1.2.5 Charger Stand
Input 100 to 240 VAC, 50Hz
/60Hz, 0.4A - 0.15A
Output 6.0VDC, 0.8 A
A1.3 Parameters
Measurement Range
SpO
2
0 % ~ 100 %
PR 25 bpm ~ 300 bpm
Alarm Range
SpO
2
0 % ~ 100 %
PR 0 bpm ~ 300 bpm
SpO
2
Accuracy
Saturation
Adult and Pediatric ± 2% (70 %~ 100 %)
Undefined (0 %~ 69 %)
Neonate ± 3% (70 %~ 100 %)
Undefined (0 %~ 69 %)