H100B Pulse Oximeter User Manual
- 80 -
A2.2 Electromagnetic Immunity - for all
EQUIPMENT and SYSTEMS
Guidance and manufacturers declaration – electromagnetic
immunity
The oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the
oximeter should assure that it is used in such an environment.
Emissions test Compliance Compliance
level
Electromagneti
c environment
- guidance
Electrostatic
discharge(ESD)
IEC/EN61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should
be wood,
concrete or
ceramic tile. If
floors are
covered with
synthetic
material, the
relative
humidity should
be at least 30%.