H100B Pulse Oximeter User Manual
- 81 -
Electrical Fast
Transient/Burst
IEC/EN61000-4-4
±2kV for
power supply
lines
±1kV for
input/output
lines (>3m)
N/A
Surge
IEC/EN61000-4-5
line to line
line to ground
N/A
Mains power
quality
should be that
of a typical
commercial or
hospital
environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC/EN61000-4-1
1
<5%UT(>95
% dip in
UT)for 0.5
cycle
40%UT(60%
dip in UT)for
5 cycles
70%UT(30%
dip in UT)for
25 cycles
<5%UT(>95
N/A
Mains power
quality
should be that
of a typical
commercial or
hospital
environment. If
the user
of the product
requires
continued
operation
during power
mains

Table of Contents