EasyManua.ls Logo

EDAN H100B - Pulse Rate Alarm Limit Configuration

EDAN H100B
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
H100B Pulse Oximeter User Manual
- 46 -
limit of SpO
2
Alarm - 1”. If the value of SpO2
Lo Alarm is set to less than 85, it will restore
to 85 after the oximeter is turned on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
SpO
2
Hi Alarm is always higher than SpO
2
Lo Alarm by at least 1%.
Press the Function button, confirm the alarm
range setup.
Press the On/Off button to exit the menu, and
return to measurement state.
3.3.11 PR Alarm Setup
The user can use PR Hi Alarm and PR Lo Alarm in
menu to adjust pulse rate alarm limits.
By default, PR Hi Alarm and PR Lo Alarm in Neo
mode are 200 and 100 respectively; while they are 120
and 50 in Adu mode respectively.
Set the PR limits as follows:
Choose PR Hi Alarm in the menu, press the
Function button to enter setup. The PR Hi
Alarm box changes from real line to broken
line. The adjustable range of the upper limit of

Table of Contents

Related product manuals