STP-iX455/iXL455 Series Instruction Manual
9-46
Operation
In case of time flag = 0
Parameter
Item Data format Remarks
a Error Code
8-bits hexadecimal
coded ASCII
*1
b Time flag
8-bits hexadecimal
coded ASCII
c Pump running time
32-bits hexadecimal
coded ASCII
Unit: minute
d Ctrl running time
32-bits hexadecimal
coded ASCII
Unit: minute
*1:
Value corresponding to the error is transmitted (see Table 9.9).
When the number of errors being detected is under the maximum number, the value of
parameter that is larger than the number of errors being detected is set to FF
hex
.
Example: In case of Disturbance X_H detection at
Pump running time = 5,000 minutes, Ctrl running time = 6,000 minutes
Parameter
a b c d
ASCII 0
D
0
0
0
0
0
0
1
3
8
8
0
0
0
0
1
7
7
0
HEX 30
44
30
30
30
30
30
30
31
33
38
38
30
30
30
30
31
37
37
30
In case of time flag = 1
Parameter
Item Data format Remarks
a Error Code
8-bits hexadecimal
coded ASCII
*1
b Time flag
8-bits hexadecimal
coded ASCII
c
Error detection time
(Formatyymmddhhnn)
yy: The last two digits of
the year
mm: Month
dd: Day
hh: Hour (24-hour display)
nn: Minute
40-bits hexadecimal
coded ASCII
*2
d [System reservation]
24-bits hexadecimal
coded ASCII
*1:
Value corresponding to the error is transmitted (see Table 9.9).
When the number of errors being detected is under the maximum number, the value of
parameter that is larger than the number of errors being detected is set to FF
hex
.
*2:
Each value of time is transmitted as a character string.
Example: In case of Disturbance X_H detection at July 15, 2006 12: 34
Parameter
A b c d
ASCII 0
D
0
1
0
6
0
7
1
5
1
2
3
4
*3
*3
*3
*3
*3
*3
HEX 30
44
30
31
30
36
30
37
31
35
31
32
33
34
*3:
System reservation