EasyManua.ls Logo

Edwards STP-iXL455

Edwards STP-iXL455
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STP-iX455/iXL455 Series Instruction Manual
4-23
Installation
1
2
3
4
5
6
7
Table 4.6 Pin Assignment for Connector X1
Connector X1
pin No
Signal Remarks Outline
1 +48 V For +48 VDC power supply
2 0 VA
For t
he ground for
DC
power supply
3, 4, 5, 6
DO NOT connect these
pins
*1
7 FG Frame ground
Connect the external power supply unit to No.1 and No.2 pins of the connector X1.
Failure to do so may result in product damage.
Refer to the iPS-240 Instruction Manual for the operation of the power supply unit
(optional accessory).
No. 2 pin (0VA) and No. 7 pin (FG) of the connector X1 are connected internally to
the pump body and other metallic components.
The STP pump mounts 7 A fuse in DC power supply input line as overcurrent
protection.
*1
: It is equipped for a power supply unit iPS-240 (optional accessory).
X1 DC POWER

Table of Contents

Related product manuals