STP-iX455/iXL455 Series Instruction Manual
13-8
Troubleshooting
13.3.1 Cleaning and Decontamination
The method of cleaning and decontamination of the STP pump is shown below.
Table 13.4 Cleaning and Decontamination
Area Cleaning and Decontamination
STP pump
Exterior
Wipe with proper solvent (such as alcohol).
When the label of the STP pump has been damaged, contact
Service office.
Interior
The decontamination (overhaul) in the STP pump is executed in
Service office. When overhaul is needed, contact Service office.
Control unit Exterior
Clean the outside of the control unit with a dry wipe as needed.
When dust has accumulated in the ventilation port, wipe off or
vacuum it with the cleaner. In this case, dust must not enter in
the control unit.
Clean off with a pure gas when dust has accumulated in the
connector.
When the label of the control unit has been damaged, contact
Service office.
Interior The decontamination (overhaul) in the control unit is executed in
Service office. When overhaul is needed, contact Service office.
13.3.2 Inspecting the Deposit
Leaving the STP pump without removing the deposit may cause
the STP pump to be corroded beyond repair.
Deposit may have accumulated inside the STP pump depending upon the type of the
vacuum equipment installed in the STP pump. An increase in the amount of deposit may
lead to a malfunction of the STP pump.
Therefore, perform regular maintenance (once every three to six months as a target).
Inspect the inside of the outlet port flange for adhesion of deposit while watching the
outlet port flange. If deposit has accumulated to some extent, an overhaul (cleaning) is
required. Contact Service office.
Note that the costs of troubleshooting problems resulting from deposit will be at your
own charge even during the warranty period.
If a problem resulting from deposit occurs inside the STP pump, hermetically seal inlet
port, outlet port, or purge port, immediately return them to Service office. For the
transport method, see Section13.4.4, "Transport for Repair or Overhaul".
".
◇ Contact with atmospheric air may cause a reaction of the deposit. DO NOT open
the STP pump to atmospheric air as much as possible.