STP-iX455/iXL455 Series Instruction Manual
9-22
Operation
9.4.2 ReadMeas
Function:
Reads the measured rotational speed.
Transmission Frame:
PCSIM
Stx
0
0
1
? D
Etx
LRC
SIMPC
Ack
PCSIM
Ack
SIMPC
Stx
0
0
1
Sp
D
Parameter 1 to 2
Etx
LRC
Parameter
Item Data format Remarks
1 [System reservation]
56-bits hexadecimal
coded ASCII
2
Measured rotational speed
(Unit: Hz)
16-bits hexadecimal
coded ASCII
Example:
Measured rotational speed: 0395
hex
= 917 Hz = 55,020 rpm
Parameter
1 2
ASCII
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
0
3
9
5
HEX
30
33
39
35
*1: