STP-iX455/iXL455 Series Instruction Manual
7-7
Operation
(4)
(16)
(5)
(17)
(18)
(19)
(6)
(21)
(8)
(9)
(20)
(7)
(10)
(22)
(11)
(23)
X2 :
Remote Output Signal
N.O OUT:
Normal Open Output Pin
N.C OUT:
Normal Close Output Pin
COM : Common Pin
CR1
CR2
CR3
CR3
CR4
CR5
CR5
CR6
CR7
POWER
N.O OUT
ACCELERATION
N.O OUT
N.O OUT
N.C OUT
NORMAL COM
N.O OUT
N.O OUT
N.O OUT
N.C OUT
COM
N.O OUT
WARNING
FAILURE
BRAKE
I/O ENABLE
POWER ON state
signal output pins
ACCELERATION state
signal output pins
NORMAL OPERATION state
signal output pins
BRAKE state
signal output pins
FAILURE state
signal output pins
WARNING state
signal output pins
PARALLEL PORT SELECT state
signal output pins
Figure 7.3 X2 REMOTE Output Signal Pins
Table 7.5 shows the rated contacts for relays CR1 to CR7 in Figure 7.3.
Table 7.5 Rated Contacts for Relays CR1 to CR7
Resistance Load (COSφ=1)
Rated Load
125 V AC, 0.3 A
30 V DC, 1 A
Rated Current 1 A
Maximum Contact Point Current
1 A
Maximum Open/Close Capacity
AC: 37.5 VA
DC: 30 W
Minimum Applicable Load 10 mV DC, 10 μA