EasyManua.ls Logo

Edwards T-STATION 85

Edwards T-STATION 85
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 9
Turbo setpoint number entry mode
Turbo setpoint screen key acons
Table 8
Turbo setpoint screen key acons
Keys Short press Long press
START Go to turbo screen -
ENTER Enter edit mode, or conrm
selecon
Enter number entry mode
NEXT Next TSP opon, next number or go
to gauge calibraon screen
Scroll through TSP opons in edit
mode, scroll through numbers in
number entry or scroll through
menu screens
Calibrate gauge screen
When the calibrate gauge screen is selected, the gauge LED and the CAL LED are lit together.
The numeric display is blank. The calibrate gauge menu is not available for gauges which do not
support calibraon.
When the ENTER key is pressed the acon depends on gauge type:
WR
G or APGX: The calibraon command is sent to the gauge and the display shows "CALd" for
3 seconds.
ASG: The calibraon funcons as a zero oset adjustment. The pressure currently displayed is
saved as the zero oset and is subtracted from all future readings. The display shows "CALd"
for 3 seconds to conrm the acon. The oset adjustment can be cancelled by pressing the
ENTER key again. The display shows "OFF" for 3 seconds to conrm that the oset adjustment
has been removed.
D395-94-880A - Operaon