Technical Reference
Note:
The
operang, storage condions and performance of the T-Staon 85 may depend on the
type of backing pump ed to the T-Staon 85; refer to the technical data in the appropriate
supplementary publicaons as listed in General descripon on page 10.
Operang and storage condions
Table 14
Operang and storage condions
Parameter Value Units
Ambient
operang temperature range 12 to +40 °C
Ambient storage temperature range -30 to +70 °C
Maximum operang humidity maximum 90
condensing at 40
% RH and °C
Maximum operang altude 2000 maximum m
Sound level 56 dB(A)
Polluon degree (EN61010) 2 -
Maximum outlet pressure
- XDD1 pump
- E2M1.5 pump
1.1 (1.1 x 10
5
)
1.5 (1.5 x 10
5
)
bar absolute (Pa)
Maximum permied external magnec eld
(nEXT85 turbo)
5 mT
Mechanical data
Table 15
Mechanical data
Parameter Data
Dimensions Refer to Figure 12
on page 37
Degree of protecon (to IEC34-5:1981) IP20
Mass:
- XDD1 combinaon
- E2M1.5 combinaon
17 kg max
21 kg max
D395-94-880A - Technical Reference
Page 36