Inspect the hoses, pipelines and connecons
1. Inspect all of the electrical connecons and check that they are secure. Tighten any
loose connecons.
2. Inspect all of the electrical cables and check that they are not damaged and have not
o
verheated. Replace or repair any damaged or overheated cable.
3. Inspect all of the vacuum and exhaust connecons and check that they are secure.
Tighten any loose connecons.
4. Inspect all of the vacuum and exhaust pipelines and check that they are not corroded
or damaged and that they do not leak. Replace or repair any corroded or damaged
component and seal any leaks found.
Factory default sengs
Procedure to return the T-Staon 85 to the factory default sengs.
1. Remove the mains power from the
T-Staon 85.
2. Press the NEXT key and hold it down whilst reapplying mains power. Connue holding
down the NEXT key unl "Err01" is displayed.
3. Remove and reapply mains power. The factory defaults will now be set.
D395-94-880A - Maintenance