Inspect the hoses, pipelines and connecons
1. Inspect all of the electrical connecons and check that they are secure. Tighten any
loose connecons.
2. Inspect all of the electrical cables and check that they are not damaged and have not
o
verheated. Replace or repair any damaged or overheated cable.
3. Inspect all of the vacuum and exhaust connecons and check that they are secure.
Tighten any loose connecons.
4. Inspect all of the vacuum and exhaust pipelines and check that they are not corroded
or damaged and that they do not leak. Replace or repair any corroded or damaged
component and seal any leaks found.
Factory default sengs
Procedure to return the T-Staon 85 to the factory default sengs.
1. Remove the mains power from the
T-Staon 85.
2. Press the NEXT key and hold it down whilst reapplying mains power. Connue holding
down the NEXT key unl "Err01" is displayed.
3. Remove and reapply mains power. The factory defaults will now be set.