Figure 15
Pin connecons for an 8-way RJ45
Table 21
Acve gauge connector pin-out
Pin Funcon
1 Power supply posive
2 Power supply common
3 Signal input
4 Idencaon
5 Signal common
6 Control line 1
7 Control line 2
8 Not connected
Electrical data
If the T-Staon 85 uses an E2M1.5, the motor start up current is drawn for less than one
sec
ond, so slow-blow fuses must be used to prevent unnecessary fuse failure when the pump
starts. If using the pump at temperatures lower than 12 °C (53.6 °F), the start up current will be
drawn for longer; this may cause the motor thermal overload device to open.
Table 22
Electrical data
Supply Voltage
50/60 Hz
Current (A) Power (W)
Full load Start
XDD1 / nEXT85
100 3.1 - 310
120 2.6 - 310
200 1.6 - 310
230 1.35 - 310
E2M1.5 / nEXT85
100 4.7 13.6 470
120 3.92 12.0 470
200 2.2 6.4 430
230 1.9 5.7 430