26
TROUBLESHOOTING
R
9 TROUBLESHOOTING
The instrument does not operate
Check:
the connection to the mains electricity supply
the power on/off switch position at the rear of the Viscometer head
The spindle does not rotate concentrically
Check:
that the spindle is correctly adjusted to the shaft
that the union between spindle and shaft is clean
The instrument does not read zero in vacuum
Check:
that the instrument is levelled correctly
Viscosity reading is not stable or not very accurate
Check:
that the instrument is levelled correctly
that the rotation speed/spindle type selection is correct
that the temperature of the sample is stable
Viscometer head slips down the column
tighten the two nylon grub screws on the head mounting bracket using the hexagonal wrench
supplied