TROUBLESHOOTING
R
9 TROUBLESHOOTING
The instrument does not operate
Check:
• the connection to the mains electricity supply
• the power on/off switch position at the rear of the Viscometer head
The spindle does not rotate concentrically
Check:
• that the spindle is correctly adjusted to the shaft
• that the union between spindle and shaft is clean
The instrument does not read zero in vacuum
Check:
• that the instrument is levelled correctly
Viscosity reading is not stable or not very accurate
Check:
• that the instrument is levelled correctly
• that the rotation speed/spindle type selection is correct
• that the temperature of the sample is stable
Viscometer head slips down the column
• tighten the two nylon grub screws on the head mounting bracket using the hexagonal wrench
supplied