55
* Contact Elcometer or your local Elcometer Supplier to arrange return.
Figure 34. Example error message - no probe is connected to gauge
UNSTABLE
READING
a) External electro-magnetic interference.
b) Gauge faulty.
a) Isolate from electro-magnetic
interference.
b) Return to Elcometer*.
VALUE TOO
LARGE
Numerical error. Switch gauge off then on again.
If error persists contact Elcometer.
CALIBRATION Incorrect calibration calculation. Re-calibrate.
If error persists, return to Elcometer*.
LANGUAGE
MEMORY
Software error. Return to Elcometer*.
Error message Causes Action to take