6 Configure the 3410 Series Electronics
1. Check all settings and wiring carefully.
2. Set up the Meter Directory, and start initial communication with the 3410 Series
electronics.
The basic instructions for setting up the directory and initializing communications are
contained in the following text. Refer to the 3410 Series Gas Ultrasonic Flow Meter
Installation Manual (P/N 3-9000-759, Section 4) for more detailed instructions. Use Daniel
MeterLink v1.10 or later to make a connection to the meter.
6.1 Initial communication connection using
Ethernet
6.1.1 Ethernet initial connection material checklist
The following materials and information are required:
• 3410 Ethernet adapter cable (Daniel P/N 1-360-01-596)
• Personal computer (PC) configured as follows:
— Daniel MeterLink software installed (version 1.10 or later)
— Ethernet LAN adapter
— Configured to automatically obtain IP address (via DHCP)
• Desired 3400 Series 3410 Series meter(s) communication configuration parameters:
— IP address
— Serial communication parameters such as baud rate and Modbus ID (if desired)
6.1.2
Ethernet initial connection steps
Procedure
1. Power up the 3410 Series meter and wait two minutes for the meter to startup.
2. Ensure that the Ethernet DHCP server is enabled on the 3410 CPU Module by
setting DHCP switch to the ON position.
3. Plug the Ethernet adapter cable (P/N 1-360-01-596) into the meter CPU Module and
connect the RJ45 end into the PC Ethernet connector.
4. Verify the Ethernet connection status by the 3410 CPU Module LINK LED which
should be solid green.
5. Start Daniel MeterLink
™
on the PC. The Connect to Meter dialog displays.
a) Click the Edit Meter Directory button.
Upgrade kit instructions Configure the 3410 Series Electronics
P/N 3-9000-784 July 2019
3410 Series Electronics Upgrade kit instructions 47