b) Click the Add button (below the Meter Directory file operations field). Daniel
MeterLink inserts the New Meter record in the last field in the Meter
Directory table.
6. Create a new meter directory record with the following parameters:
a) Enter a name for the meter.
b) Select Meter Type as Gas or Liquid as appropriate.
c) Check the Ethernet box and uncheck the Direct and Modem boxes.
d) Set the Ethernet IP address to 192.168.135.100.
e) Close the Meter Directory window.
7. Use the Meter|Connect menu path and click the Ethernet connection for the
meter record established in the previous step. Daniel MeterLink
will now connect
to the meter using the user-specified Ethernet settings.
6.2 Initial communication connection using RS-232
Serial
6.2.1 Communications settings and Field Setup Wizard
RS-232 Serial initial connection material checklist
The following materials and information are required:
Ultrasonic serial cable (Daniel P/N 3-2500-401).
Personal computer (PC) with Daniel MeterLink
software version 1.10 (or later)
Desired 34003410 meter(s) communication configuration parameters:
IP address (if desired)
Serial communication parameters such as baud rate and Modbus ID
RS-232 Serial initial connection steps
Procedure
1. Power up the 3410 meter and PC.
2. Plug the DB-9 end of the cable directly into the PC running Daniel MeterLink. The
three wires on the other end of the cable connect to RX, TX, and COM terminals on
the CPU Module. The Digital Out wire goes to RX, the Digital In wire goes to TX, and
the COMMON wire goes to COM.
Configure the 3410 Series Electronics Upgrade kit instructions
July 2019 P/N 3-9000-784
48 3410 Series Electronics Upgrade kit instructions