• Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow,
lags Phase A while reporting reverse flow)
Phase A and Phase B output (based on flow direction):
• Reverse flow - output only reports flow in the reverse direction. For frequency
outputs, Phase B of the output is 90 degrees out of phase with Phase A.
• Forward flow - output only reports flow in the forward direction. For frequency
outputs, Phase B of the output is 90 degrees out of phase with Phase A.
• Absolute - output reports flow in both directions. For frequency outputs, Phase B
of the output is 90 degrees out of phase with Phase A.
• Bidirectional - output reports flow on Phase A only in the forward direction and
on Phase B only in the reverse direction.
Maximum frequency for the frequency outputs:
• 1000 Hz
• 5000Hz
3. If changes were made in the Field Setup Wizard, choose to save the configuration
when Daniel MeterLink
™
prompts you after the changes are written to the meter.
Click Finish in the Field Setup Wizard dialog to write any changes to the meter.
4. You have successfully completed upgrading your3400 electronics to 3410 Series
electronics.
5. Save your configuration file for the meter's parameters you just set as the “As Left
Settings”.
Upgrade kit instructions Configure the 3410 Series Electronics
P/N 3-9000-784 July 2019
3410 Series Electronics Upgrade kit instructions 59