Configuration and Use Manual 29
Flowmeter Startup
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Figure 5-6 Communicator – mA output trim procedure
Figure 5-7 Communicator – Scaled AO trim procedure
Read mA output at
receiving device
Read mA output at
receiving device
Equal?
Enter receiving device
value at Communicator
Read mA output at
receiving device
Read mA output at
receiving device
Equal?
Enter receiving device
value at Communicator
4 mA trim 20 mA trim
Yes
No
No
Yes
ENTER ENTER
OK
OK
On-Line Menu >
3 Diag/Service >
6 Trim Analog Out 1/2
Enter low mA value
ENTER
OK
On-Line Menu >
3 Diag/Service >
7 Scaled AO Trim
CHANGE
Enter high mA value
ENTER
PROCEED
Trim at low mA value
Trim at high mA value
OK