EasyManua.ls Logo

Emerson Micro-Motion 2400S - Zero Procedure

Emerson Micro-Motion 2400S
198 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration and Use Manual 31
Flowmeter Startup
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
5.5.2 Zero procedure
To zero the flowmeter:
• Using the display menu, see Figure 5-8. For a complete illustration of the display zero menu,
see Figure C-16.
• Using the zero button, see Figure 5-9.
• Using ProLink II, see Figure 5-10.
• Using the Communicator, see Figure 5-11.
Note the following:
• If the transmitter was ordered with a display:
- The zero button is not available.
- If the off-line menu has been disabled, you will not be able to zero the transmitter with the
display. For information about enabling and disabling the off-line menu, see
Section 8.10.3.
- You cannot change the zero time with the display. If you need to change the zero time, you
must use the Communicator or ProLink II.
• If the transmitter was ordered without a display, the zero button is available.
- You cannot change the zero time with the zero button. If you need to change the zero time,
you must use the Communicator or ProLink II.
- The zero button is located on the user interface board, beneath the transmitter housing
cover (see Figure 2-2). For instructions on removing the transmitter housing cover, see
Section 2.3.
- To press the zero button, use a fine-pointed object such as the end of a paperclip. Hold the
button down until the status LED on the user interface module begins to flash yellow.
• During the zero procedure, the status LED on the user interface module flashes yellow.

Table of Contents