Mass/volume conversion example: Mass/volume conversion example
Time conversion example: Time conversion example
Mass/volume and time conversion example: Mass/volume and time conversion
example
Mass/volume conversion example
To find the Base per Custom relationship for a custom unit of kilograms per hour (kg/h)
and a base unit of grams per hour (g/h), input the following:
Custom Unit = kg/h Base Unit = g/h
Because:
1 kg (Kilogram) = 1000 g (Grams)
Then: 1 kg/h = × = 1000 g/h
1 kg/h = 1000 g/h
Therefore:
Base per Custom = = = 1000
Figure 2-28: Flow Rate Custom Units - Mass/Volume Conversion Example
Time conversion example
To find the Base per Custom relationship for a custom unit of standard cubic feet per hour
(scf/h) and a base unit of standard cubic feet per minute (StdCuft/min), input the
following:
Custom Unit = scf/h Base Unit = StdCuft/min
Because:
Configuration Reference Manual
October 2018 00809-0100-4803
46 Rosemount 3051S Multivariable Transmitter