Figure 3-8: SuperModule Isolator Plate
A. Bolt
B. SuperModule isolator plate
C. Coplanar flange
D. Flange adapters
To install adapters to a coplanar flange, perform the following procedure:
Procedure
1. Remove the flange bolts.
2. Leaving the flange in place, move the adapters into position with the O-rings
installed.
3. Attach the adapters and the coplanar flange to the transmitter SuperModule
assembly using the longer of the bolts supplied.
4. Tighten the bolts. Refer to Table 3-2 for torque specifications.
Refer to for the correct part numbers of the flange adapters and O-rings designed
for the Rosemount 3051SMV.
Note
The two styles of Rosemount flange adapters (Rosemount 3051S/3051/2051) each
require a unique O-rings. Use only the O-ring designed for the corresponding flange
adaptor.
O-rings
The two styles of Rosemount
™
flange adapters (Rosemount 1151 and Rosemount
3051/2051/2024/3095) each require a unique O-ring (see Figure 3-9). Use only the O-ring
designed for the corresponding flange adapter.
Installation
Reference Manual
October 2018 00809-0100-4803
82 Rosemount 3051S Multivariable Transmitter