Rosemount 3100 Series
4-14
NOTE:
When using the green button
to scroll through the list of measurement
units, allow three seconds after each button press for the transmitter to
check and display the selection.
NOTE:
After changing units, a scaling factor (see page 4-19) needs to be edited
to see the correct PV.
Setting the Correct
Bottom Reference
(The 3102/3105)
Screen display: b.rEF
Factory default value: 11
The transmitter leaves the factory with the bottom reference pre-programmed
to the maximum range of the instrument 36 ft (11 m).
NOTE:
It is important to not enter a value greater than the maximum range of the
transmitter, which is 36 ft (11 m).
Figure 4-3. Bottom Reference
NOTE:
This menu option is in the programming menu. See Figure C-2 on page C-3
for a map of the menu structure and how to access the menu options.
To change the bottom reference:
a) If entering the menu system from the PV display, press the green button
repeatedly until the “b.rEF” menu option is indicated (see note above.)
b) Press the blue button
to enter the “b.rEF” menu and display the
present bottom reference (b.rEF) value.
c) If this value is correct, press the red button
and then the green button to
get to the next menu option.
d) Press the blue button
to start the editing mode. The first digit flashes to
indicate it can now be edited.
e) Press the green button
repeatedly to edit the flashing digit.
f) Press the blue button
to select the next digit. The digit flashes to
indicate it can be edited.
Distance-to-target
Depth (Level)
Bottom Reference