90
Reference Manual
00809-0100-4021, Rev GC
Section 3: HART commissioning
April 2014
HART® commissioning
Active Sensor Drift Alerts
Viewing Active Sensor Drift Alerts: Fast Keys 1-1-2
When the Sensor Drift Alert diagnostic detects a drifting sensor, the LCD display will display a
message; “ALARM DRIFT ALERT” if configured in Alarm Mode and “WARN DRIFT ALERT” if
configured in Warning Mode.
1. Select 1 Overview.
2. If Sensor Drift Alert is configured in Alarm Mode, select 1 Device Status: Failed.
If Sensor Drift Alert is configured in Warning Mode, select 1 Device Status:
Maintenance.
3. Select 2 Sensor Drift Alert Active.