89
Reference Manual
00809-0100-4021, Rev GC
Section 3: HART commissioning
April 2014
HART® commissioning
To verify that Sensor Drift Alert is enabled: Fast Keys 2-2-4-2
1. From the Home Screen, select 2 Configure.
2. Select 2 Manual Setup.
3. Select 4 Diagnostics.
4. Select 2 Sensor Drift Alert.
5. You will see this screen. Under 1 Mode, it will say either Alarm or Warning if enabled, or
Disable. If enabled, it will also display the current diagnostic values.