HART® commissioning
6. Select the engineering units you would like to use and press ENTER. Choose from degC,
degF, degR, Kelvin, mV, Ohms.
7. Enter the Sensor Drift Alert threshold value and press ENTER. This is a digital value that
triggers the Drift Alert feature. When this limit is exceeded, the transmitter will go into
alarm or generate a warning (depending on the alert mode chosen previously).
8. Enter a damping value between 0 and 32 and press ENTER. This damping value is
additional damping applied to the result of (S1-S2) after each sensor's individual
damping value has already been applied.
9. Configuration is complete. Press OK.