83
Reference Manual
00809-0100-4021, Rev GC
Section 3: HART commissioning
April 2014
HART® commissioning
Disable Sensor Drift Alert in guided setup: Fast Keys 2-1-6
1. From the Home Screen, select 2 Configure.
2. Select 1 Guided Setup.
3. Select 6 Config Drift Alert.
4. Select 2 Disable to disable Sensor Drift alert and press ENTER.
5. Sensor Drift Alert has been disabled. Press OK.