Cerabar M, Deltabar M, Deltapilot M Commissioning with the FF configuration program
Endress+Hauser 219
9.12.5 Backup or duplicate device data
The device has no memory module. With an operating tool based on FDT technology (e.g.
FieldCare), you have the following options (see parameter "Download select" ä 119 in
operating menu or via Resource block ä 173):
• Saving/rescuing configuration data
• Duplicating instrument configurations
• Transferring all relevant parameters when replacing electronic inserts.
For more information, read the operating manual for the FieldCare operating program.