EasyManua.ls Logo

Endress+Hauser Ceraphant PTP31B - Relay Switching Capacity

Endress+Hauser Ceraphant PTP31B
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical connection Ceraphant PTP31B, PTP33B
20 Endress+Hauser
2 x PNP switch outputs R1 and R2
M12 plug Valve plug Cable
L–
L+
2
1
3
4
R1
R2
0.63A
A0023248
-
L–
L+
1
2b
2a
3
4
R2
(a)
0.63A
R1
"
A0023282
1 brown = L+
2a black = switch output 1
2b white = switch output 2
3 blue = L-
4 green/yellow = ground
(a) reference air hose
1 x PNP switch output R1 with additional analog output 4 to 20 mA (active)
M12 plug Valve plug Cable
L–
L+
2
1
3
4
R1
0.63A
A0023249
-
L–
L+
1
2b
2a
3
4
(a)
0.63A
R1
"
A0030519
1 brown = L+
2a black = switch output 1
2b white = analog output 4 to 20 mA
3 blue = L-
4 green/yellow = ground
(a) reference air hose
6.1.2 Supply voltage
Supply voltage: 10 to 30 V DC
6.1.3 Current consumption and alarm signal
Intrinsic power consumption Alarm current (for devices with analog output)
≤ 60 mA ≥21 mA (factory setting)
6.2 Relay switching capacity
Switch state ON: I
a
≤ 250 mA; switch state OFF: I
a
≤1 mA
Switch cycles: >10,000,000
Voltage drop PNP: ≤2 V
Overload protection: Automatic load testing of switching current;
Max. capacitance load: 14 μF at max. supply voltage (without resistive load)
Max. cycle duration: 0.5 s; min. t
on
: 4 ms
Periodic disconnection from protective circuit in the event of overcurrent (f = 2 Hz) and
"F804" display