Technical data Ceraphant PTP31B, PTP33B
68 Endress+Hauser
15 Technical data
15.1 Input
15.1.1 Measured variable
Measured process variable
Gauge pressure or absolute pressure
Calculated process variable
Pressure
15.1.2 Measuring range
Metal process isolating diaphragm
Sensor Device Maximum
Sensor measuring range
Lowest
calibratable
span
1)
MWP OPL Factory settings
2)
Option
3)
lower (LRL) upper (URL)
[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)]
Devices for gauge pressure measurement
400 mbar (6 psi)
4)
PTP31B
PTP33B
–0.4 (–6) +0.4 (+6) 0.4 (0.6) 1 (15) 1.6 (24) 0 to 400 mbar (0 to 6 psi) 1F
1 bar (15 psi)
4)
PTP31B
PTP33B
–1 (–15) +1 (+15) 1 (15) 2.7 (40.5) 4 (60) 0 to 1 bar (0 to 15 psi) 1H
2 bar (30 psi)
4)
PTP31B
PTP33B
–1 (–15) +2 (+30) 0.4 (0.6) 6.7 (100.5) 10 (150) 0 to 2 bar (0 to 30 psi) 1K
4 bar (60 psi)
4)
PTP31B
PTP33B
–1 (–15) +4 (+60) 0.8 (1.2) 10.7 (160.5) 16 (240) 0 to 4 bar (0 to 60 psi) 1M
10 bar (150 psi)
4)
PTP31B
PTP33B
–1 (–15) +10 (+150) 2 (30) 25 (375) 40 (600) 0 to 10 bar (0 to 150 psi) 1P
40 bar (600 psi)
4)
PTP31B
PTP33B
–1 (–15) +40 (+600) 8 (120) 100 (1500) 160 (2400) 0 to 40 bar (0 to 600 psi) 1S
100 bar (1 500 psi)
4)
PTP31B –1 (–15) +100 (+1500) 20 (300) 100 (1500) 160 (2400) 0 to 100 bar (0 to 1 500 psi) 1U
400 bar (6 000 psi)
4)
PTP31B –1 (–15) +400 (+6000) 80 (1200) 400 (6000) 600 (9000) 0 to 400 bar (0 to 6 000 psi) 1W