Parameter description (FF configuration program) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
64 Endress+Hauser
Unit Volume/
VOLUME_UNIT
Selection
Index: 53
Data type: unsigned16
Access: auto, OOS
Select the volume unit.
Prerequisite:
Comb. Measurand/COMBINED_MEASURAND = Height and volume or
%-height and volume
Options:
•l
•hl
•cm
3
•dm
3
•m
3
•m
3
E
3
•ft
•ft
3
E
3
•gal
•Igal
•bbl
•User unit, see also Customer Unit V/CUSTOMER_UNIT_VOLUME ( ä 64) and
Cust. Unit. Fact. V/CUSTOMER_UNIT_FACTOR_VOLUME ( ä 64)
Factory setting:
m
3
Customer Unit V/
CUSTOMER_UNIT_
VOLUME
Entry
Index: 54
Data type: visible string
Access: auto, OOS
Enter text (unit) for customer-specific volume unit.
You can enter a maximum of eight alphanumeric characters here.
See also Cust. Unit. Fact. V/CUSTOMER_UNIT_FACTOR_VOLUME.
Prerequisite:
Comb. Measurand/COMBINED_MEASURAND = Height and volume,
Unit Volume/VOLUME_UNIT = User unit
Comb. Measurand/COMBINED_MEASURAND = %-height and volume,
Unit Volume/VOLUME_UNIT = User unit
Comb. Measurand = COMBINED_MEASURAND
Unit Volume = VOLUME_UNIT
!
Note!
Only the first five characters are shown on the local operation. For example, if the customer-
specific unit "crates" was selected, only "crate" would be displayed.
If the unit contains a slash, up to eight characters can be shown on the local operation. The
maximum number of characters in the counter is again limited to five. For example if the
customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. In the
FF configuration program, the slash counts as a character, i.e. "crates/m2" would be
displayed as "crates/m".
Factory setting:
_ _ _ _ _ _ _ _
Cust. Unit. Fact. V/
CUSTOMER_UNIT_
FACTOR_VOLUME
Entry
Index: 55
Data type: float
Access: auto, OOS
Enter the conversion factor for a customer-specific volume unit.
The conversion factor must be entered in relation to the SI unit "m
3
".
See also Customer Unit V/CUSTOMER_UNIT_VOLUME.
Prerequisite:
Comb. Measurand/COMBINED_MEASURAND = Height and volume,
Unit Volume/VOLUME_UNIT = User unit
Comb. Measurand/COMBINED_MEASURAND = %-height and volume,
Unit Volume/VOLUME_UNIT = User unit
Comb. Measurand = COMBINED_MEASURAND
Unit Volume = VOLUME_UNIT
Example:
You want the measured value to be displayed in "buckets".
Primary Value/PRIMARY_VALUE = 0.01 m3 i 1 bucket
Enter Customer Unit V/CUSTOMER_UNIT_VOLUME: bucket
Enter Cust. Unit. Fact. V/CUSTOMER_UNIT_FACTOR_VOLUME : 100
Result: Primary Value/PRIMARY_VALUE = 1 bucket