Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Parameter description (FF configuration program)
Endress+Hauser 65
Mass Unit/
MASS_UNIT
Selection
Index: 56
Data type: unsigned16
Access: auto, OOS
Select the mass unit.
Prerequisite:
Lind. Measurand./LINEARIZED_MEASURAND = Pressure and mass
Options:
•g
•kg
•t
•oz
•lb
•ton
•User unit, see also the following parameter descriptions for Customer Unit M/
CUSTOMER_UNIT_MASS ( ä 65) and Cust. Unit Fact. M/CUSTOMER_UNIT_
FACTOR_MASS ( ä 65)
Factory setting:
kg
Customer Unit M/
CUSTOMER_UNIT_
MASS
Entry
Index: 57
Data type: visible string
Access: auto, OOS
Enter text (unit) for customer-specific mass unit.
You can enter a maximum of eight alphanumeric characters here.
See also Cust. Unit Fact. M/CUSTOMER_UNIT_FACTOR_MASS.
Prerequisite:
Lind. Measurand./LINEARIZED_MEASURAND = Pressure and mass,
Mass Unit/MASS_UNIT = User unit
!
Note!
Only the first five characters are shown on the local operation. For example, if the customer-
specific unit "crates" was selected, only "crate" would be displayed.
If the unit contains a slash, up to eight characters can be shown on the local operation. The
maximum number of characters in the counter is again limited to five. For example if the
customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. In the
FF configuration program, the slash counts as a character, i.e. "crates/m2" would be
displayed as "crates/m".
Factory setting:
_ _ _ _ _ _ _ _
Cust. Unit Fact. M/
CUSTOMER_UNIT_
FACTOR_MASS
Entry
Index: 58
Data type: float
Access: auto, OOS
Enter the conversion factor for a customer-specific mass unit.
The conversion factor must be entered in relation to the SI unit "kg".
See also Customer Unit M/CUSTOMER_UNIT_MASS.
Prerequisite:
Lind. Measurand./LINEARIZED_MEASURAND = Pressure and mass,
Mass Unit/MASS_UNIT = User unit
Example:
You want the measured value to be displayed in "buckets".
Primary Value/PRIMARY_VALUE = 10 kg i 1 bucket
Enter Customer Unit M/CUSTOMER_UNIT_MASS: bucket
Enter Cust. Unit Fact. M/CUSTOMER_UNIT_FACTOR_MASS: 0.1
Result: Primary Value/PRIMARY_VALUE = 1 bucket
Factory setting:
1.0
Pressure Transducer Block