Parameter description (FF configuration program) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
80 Endress+Hauser
Sensor Changes/
SENSOR_CHANGES
Display
Index: 42
Data type: float
Access: read only
Displays the number of sensor changes that have been performed to date.
P Peakhold Step/
PRESSURE_PEAK_
HOLD_STEP
Display
Index: 43
Data type: float
Access: read only
Displays the step size at which the pressure peakhold values are saved. The value is displayed
as a factor of the sensor end value. The pressure peakhold values are saved every 15 minutes
regardless of this value.
T. Peakhold Step/
TEMP_PEAK_HOLD_
STEP
Display
Index: 44
Data type: float
Access: read only
Displays the step size at which the temperature peakhold values are saved. The value is
displayed in the Kelvin unit. The temperature peakhold values are saved every 15 minutes
regardless of this value.
Acc. Of Gravity/
ACCELERATION_OF_
GRAVITY
Display
Index: 45
Data type: float
Access: read only
Displays the gravitational acceleration used by the device to calculate the measured variables
in the "Level" measuring mode.
Creep Flow Hyst/
CREEP_FLOW_HYST
Display
Index: 46
Data type: float
Access: read only
Displays the hysteresis of the low flow cut off. ä 80, Pressure Transducer Block,
parameter description for Set. L. Fl. Cut-Off/SET_LOW_FLOW_CUT_OFF.
Factory setting:
1 % (of end flow value)
Hist. Saving Cycle/
HISTOROM_SAVING_
CYCLE_TIME
Display
Index: 47
Data type: unsigned8
Access: read only
Displays the time interval for saving the measured value.
HistoROM Avail./
HISTOROM_AVAILABLE
Display
Index: 48
Data type: unsigned8
Access: read only
Indicates whether the optional HistoROM®/M-DAT memory module is connected to the
electronic insert.
See also Operating Instructions BA00301P (Deltabar S) or BA00302P (Cerabar S) or
BA00372P (Deltapilot S), "HistoROM
®
/M-DAT (optional)" section.
Options:
•Yes (HistoROM
®
/M-DAT is attached to the electronic insert)
No (HistoROM
®
/M-DAT is not attached to the electronic insert)