EasyManua.ls Logo

Endress+Hauser Gammapilot M FMG60 - Installation conditions

Endress+Hauser Gammapilot M FMG60
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gammapilot M FMG60 PROFIBUS PA Installation
Endress+Hauser 15
A
1
B
C
A0018076
A Vertical beam (90°)
B Diagonal beam (30°)
C Measurement section
1 Sample point
To increase the accuracy of density measurements, the use of a collimator is
recommended. The collimator screens the detector against environmental radiation.
When planning, the total weight of the measuring system must be taken into
consideration.
The Gammapilot M should be given additional support to secure it against falling or to
prevent damage to the connecting cable.
A clamping device (FHG61) and a measurement section (FHG62) are available as
accessories (Operating Instructions -> "Accessories" section)
3.3 Water cooling
See the Operating Instructions
BA00236F/00/EN
3.4 Post-installation check
After installing the measuring device, carry out the following checks:
Is the device damaged (visual inspection)?
Does the device match the measuring point specifications (ambient temperature, measuring
range etc.)?
If available: are the measuring point number and labeling correct (visual inspection)?
Is the measuring device sufficiently protected against sunlight?
Are the cable glands tightened correctly?