Gammapilot M FMG60 PROFIBUS PA Wiring
Endress+Hauser 23
1. Using a 3 mm Allen key, loosen the cover clamp for the connection compartment cover
2. Unscrew the cover
3. Push the power cable and (if required) the signal cable through the appropriate cable
glands or cable entries
4. Wire up according to the terminal assignment diagram
5. Tighten the cable glands or cable entries so that they are leak-tight
6. Screw the cover securely back onto the connection compartment
7. Turn the cover clamp so that it is set over the cover and then tighten.
4.8 Wiring in connection compartment 2
L
CAUTION
Note the following before connecting:
Connect the potential matching line to the external ground terminal of the transmitter and
to the ground terminal of the water cooling jacket (if present) before connecting the device.
(See "Potential equalization" section)
The cables must be adequately insulated, with due consideration given to the supply
voltage and the overvoltage category.
The connecting cables must offer adequate temperature stability, with due consideration
given to the ambient temperature.