Proline Promag 50H, 53H
44 Endress+Hauser
Process connections orderable only as accessories with O-ring seal, DN 2 to 25 / ¹₁₂ to 1"
Coupling SMS 1145,
threaded adapter
1.4404 (316L)
Sensor
DN
Fits to
piping
SMS
1145 DN
di G L H × B
5*H**-5*********** [mm] OD [mm] [mm] [mm] [mm] [mm]
A0005564
25
(1" ANSI)
1" 25 22.6 Rd 40 × 1/6" 30.8 72 × 55
Fitting length = (2 × L) + 86 mm
Surface roughness: R
a
0.8 m
If pigs are used for cleaning, it is essential to take the inside diameters
of measuring tube and process connection (di) into account!
External thread
1.4404 (316L)
Sensor DN Fits to
internal
thread
di G L S H × B
DKH**-GD** [mm] NP [mm] [inch] [mm] [mm] [mm]
A0005563
2 to 8 NPT ³"10³" 50 15.5 62 × 42
15 NPT ½" 16 ½" 50 20.0 62 × 42
25 (1" ANSI) NPT 1" 25 1" 55 25.0 72 × 55
Fitting length = (2 × L) + 86 mm
Surface roughness: R
a
1.6 m
Internal thread
1.4404 (316L)
Sensor
DN
Fits to
external
thread
di G D L S H × B
DKH**-GC** [mm] NP [mm] [inch] [mm] [mm] [mm] [mm]
A0005565
2 to 8 NPT ³"8.9³" 22 45 13 62 × 42
15 NPT ½" 16.0 ½" 27 45 14 62 × 42
25
(1" ANSI)
NPT 1" 27.2 1" 40 51 17 72 × 55
Fitting length = (2 × L) + 86 mm
Surface roughness: R
a
1.6 m
G
di
L
H × B
G
di
L
H × B
S
D
G
di
L
H × B
S