Device Functions Proline Promag 55 FOUNDATION Fieldbus 4 Block QUICK SETUP
Endress+Hauser 17
! Note!
The display returns to the cell SETUP COMMISSIONING (1002) if you press the X key combination during parameter
interrogation. The stored parameters remain valid.
The system units selected via the Quick Setup only apply for displaying on the local display and for parameters in the
Transducer Blocks. They do not affect the process variables (volume flow, calculated mass flow, totalizer) that are
transmitted via FOUNDATION Fieldbus.
m Only units not yet configured in the current Setup are offered for selection in each cycle. The unit for mass and
volume is derived from the corresponding flow unit.
n The YES option remains visible until all the units have been configured.
NO is the only option displayed when no further units are available.
o The "automatic parameterization of the display" option contains the following basic settings/factory settings:
YES Main line = Volume flow
Additional line = Totalizer 1
Information line = Operating/system condition
NO The existing (selected) settings remain.