EasyManua.ls Logo

Endress+Hauser Proline Promag 55 - Function Group OPERATION

Endress+Hauser Proline Promag 55
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Device Functions Proline Promag 55 FOUNDATION Fieldbus 5 Block USER INTERFACE
Endress+Hauser 23
5.1.3 Function group OPERATION
USER INTERFACE C CONTROL CAA BASIC CONFIGURATION 200
UNLOCKING/LOCKING 202
OPERATION 204
Function description
USER INTERFACE CONTROL OPERATION
TEST DISPLAY
(2040)
Use this function to test the operability of the local display and its pixels.
Options:
OFF
ON
Factory setting:
OFF
Test sequence:
1. Start the test by selecting ON.
2. All pixels of the main line, additional line and information line are darkened for
minimum 0.75 seconds.
3. Main line, additional line and information line show an "8" in each field for minimum
0.75 seconds.
4. Main line, additional line and information line show a "0" in each field for minimum
0.75 seconds.
5. Main line, additional line and information line show nothing (blank display) for
minimum 0.75 seconds.
When the test completes the local display returns to its initial state and the setting
changes to OFF.

Table of Contents

Other manuals for Endress+Hauser Proline Promag 55

Related product manuals