Device Functions Proline Promag 55 FOUNDATION Fieldbus 7 Block BASIC FUNCTION
Endress+Hauser 51
7.4 Group SENSOR DATA
7.4.1 Function group CONFIGURATION
BASIC FUNCTION G FOUNDATION FIELDBUS GGA
PROCESS PARAMETER GIA
SYSTEM PARAMETER GLA
SENSOR DATA GNA CONFIGURATION 680
Function description
BASIC FUNCTION SENSOR DATA CONFIGURATION
All sensor data (calibration factors, zero (point) and nominal diameter) are set at the factory and saved on the S-DAT
sensor memory chip.
! Note!
The individual values of the functions are also provided on the sensor nameplate.
" Caution!
Under normal circumstances you should not change the following parameter settings, because changes affect numerous
functions of the entire measuring facility in general and the accuracy of the measuring system in particular. For this
reason, the functions described below cannot be changed even when you enter your private code.
Contact the Endress+Hauser service organization if you have any questions about these functions.
K-FACTOR POSITIVE
(6801)
Use this function to display the actual calibration factor (positive flow direction) for the
sensor. The calibration factor is determined and set at the factory.
User interface:
5-digit fixed point number: 0.5000 to 2.0000
Factory setting:
Depends on nominal diameter and calibration
K-FACTOR NEGATIVE
(6802)
Use this function to display the actual calibration factor (negative flow direction) for the
sensor. The calibration factor is determined and set at the factory.
User interface:
5-digit fixed point number: 0.5000 to 2.0000
Factory setting:
Depends on nominal diameter and calibration
ZERO POINT
(6803)
Use this function to view the actual zero-point correction value for the sensor.
Zero-point correction is determined and set at the factory.
User interface:
Max. 4-digit number: –1000 to +1000
Factory setting:
Depends on nominal diameter and calibration
NOMINAL DIAMETER
(6804)
Use this function to view the nominal diameter of the sensor. The nominal diameter
depends on the size of the sensor and is set at the factory.
User interface:
2 to 2000 mm or 1/12 to 78"
Factory setting:
Depends on the size of the sensor