Device Functions Proline Promag 55 FOUNDATION Fieldbus 6 Block TOTALIZER
Endress+Hauser 39
6.2 Group HANDLING TOTALIZER
TOTALIZER D ⇒ TOTALIZER 1 DAA
⇓
TOTALIZER 2 DAB
⇓
TOTALIZER 3 DAC
⇓
HANDLING TOTALIZER DJA ⇒ Handling totalizer functions
Function description
TOTALIZER → HANDLING TOTALIZER→ Handling totalizer functions
RESET ALL TOTALIZERS
(3800)
Use this function to reset the totals (including all overflows) of the Totalizers (1 to 3) to
zero (= RESET).
Options:
NO
YES
Factory setting:
NO
FAILSAFE ALL TOT
(3801)
Use this function to define the common response of all Totalizers (1 to 3) in case of error.
Options:
STOP
The totalizer is paused until the fault is rectified.
ACTUAL VALUE
The totalizers continue to count based on the actual flow measured value. The fault is
ignored .
HOLD VALUE
The totalizers continue to count the flow based on the last valid flow value (before the
fault occurred).
Factory setting:
STOP