8 Block SUPERVISION Device Functions Proline Promag 55 FOUNDATION Fieldbus
56 Endress+Hauser
VALUE SIMULATION
MEASURAND
(8044)
! Note!
The function is not visible unless the SIMULATION MEASURAND function (8043) is
active.
Use this function to specify a selectable value (e.g. 12 m
3
/s). This is used to test the
associated functions in the device itself and downstream signal loops.
User input:
5-digit floating-point number [unit]
Factory setting:
0 [unit]
" Caution!
The setting is not saved if the power supply fails.
The appropriate unit is taken from the function group SYSTEM UNITS (ACA),
( Page 11).
SYSTEM RESET
(8046)
Use this function to perform a reset of the measuring system.
Options:
NO
RESTART SYSTEM (restart without interrupting power supply)
Factory setting:
NO
OPERATION HOURS
(8048)
Use this function to view the hours of operation of the device.
User interface:
Depends on the number of hours of operation elapsed:
Hours of operation <10 hours
display format = 0:00:00 (hr:min:sec)
Hours of operation 10 to 10000 hours
display format = 0000:00 (hr:min)
Hours of operation >10 000 hours
display format = 000000 (hr)
Function description
SUPERVISION SYSTEM OPERATION